PET - Intrinsic Viscosity - ASTM D4603

The PSL Rheotek RPV-2 determins the Inherent Viscosity of Poly (Ethylene Terephthalate)/PET in dilute solution using glass capillary viscometers in accordance with ASTM D4603.

The test method requires a 0.5% concentration of PET dissolved in 60/40 phenol/1,1,2,2-tetrachloroethane solution. Solvent and solution flow times are measured at 30 deg. C using the same 1B size of suspended-level Ubbelohde type viscometer for both measurements.

The method requires that three successive flow times are obtained within 0.2s of the mean. The RPV-2 software automatically calculates the Intrinsic Viscosity from each single determination using the Billmeyer equation.

The dual position RPV-2 (2) is an ideal configuration as well as the four position RPV-2 (4). The iSP-1 Sample Preparation system can be used in conjunction with the RPV-2 to prepare samples on a precise weight to weight basis. Dissolution is carried out using a reaction block with stirring and heating at 110 deg. C with a programmable timer. (SD-BM).

1 item(s)

  • RPV-2 Polymer Viscometer

    RPV-2 Polymer Viscometer PSL-Rheotek

    RPV- 2 (2) Request Quotation

    RPV-2 (4) Request Quotation

    • Intrinsic Viscosity
    • K-Value
    • Limiting Viscosity
    • Viscosity Number
    • Inherent Viscosity
    • Relative Viscosity

    The PSL-Rheotek RPV-2 Polymer Viscometer is suitable for measuring the dilute solution viscosity of a wide range of polymers. Applications include PA, PET, PVC, PLA, PBT, PC, PE, PP, Cellulose Acetate, Inks, Biopolymers, Pulp and cellulose.

    RPV-2  is designed for full integration with sample preparation and sample dissolution modules from PSL-Rheotek. Using these integrated options ensures full integrity of data from end to end in the measurement process.